10.1 Batch reprocessing queue file handling
IC Net 2.1
199
10 Batch reprocessing
Reprocessing is understood to be the subsequent reprocessing
of a series of chromatograms which have been loaded in a
queue (Batch reprocessing queue). For reprocessing accord-
ing to a selected method the settings for calibration, integration,
passport, appearance and report can be altered at will.
A batch reprocessing queue is stored in a batch reprocessing
queue file
*.bar
in the
Data
directory.
10.1 Batch reprocessing queue file handling
10.1.1 Open batch reprocessing queue
IC NET / File / Open / Batch reprocessing
Load an existing batch reprocessing file (
*.bar
) from the
Data
di-
rectory and open the
Reprocess
window.
IC NET / File / Open / Last batch
Load the last opened batch reprocessing file (
*.bar
) from the
Data
directory and open the
Reprocess
window.
10.1.2 Create new batch reprocessing queue
IC NET / File / Open / Chromatogram
For creation of a new batch reprocessing queue, open the
Chro-
matogram open
window. Select the desired chromatograms (
*.chw
)
and click the
<To Batch>
button. Enter the name of the new batch
reprocessing file and press
<OK>
. The selected chromatograms
are then added to this batch reprocessing queue.
If the reprocessing includes reintegration and/or recalibration,
only chromatograms recorded with the same method should be
loaded into the batch reprocessing queue.