Metrohm 758 KFD Titrino Manuel d'utilisateur Page 89

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 210
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 88
3.2 Remote control commands
758 KFD Titrino
85
*Parameter Tree part "Parameters for TIP"
.Sequence Sequence
.1 Step 1
.Select Step selection method,pause,L4 output,L6 output,
info,prep,stirrer,OFF 3.2.2.45.
.Method Method from mem.or card special 3.2.2.46.
.Pause Waiting time 0...999999, INF ditto
.L4Output Line L4 active,inactive,pulse,OFF ditto
.L6Output Line L6 active,inactive,pulse,OFF ditto
.Info Display information up to 16 ASCII char. ditto
.Prep Preparation of titrating buret internal D0, external D1,
external D2 ditto
.Stirrer Stirrer ON, OFF ditto
: up to 30 steps
.Statistics Statistics
.Status Status of statistics calculation ON, OFF 3.2.2.26.
.MeanN No. of individual determinations 2...20 ditto
.ResTab Result table
.Select original,delete n,delete all ditto
.DelN Deletion of individual results 1...20 ditto
.Presel Preselections
.IReq Request of Id's after start id1, id1&2, all, OFF 3.2.2.28.
.SReq Request of sample size after start value, unit, all, OFF ditto
.LimSmplSize Limits for sample size 3.2.2.29.
.Status Status of limit control ON, OFF ditto
.LoLim Lower limit 0.0...999 999 ditto
.UpLim Upper limit 0.0...999 999 ditto
.MeasMode Measuring mode for man.meas. pH,U,Ipol,Upol,T,OFF 3.2.2.47.
.MeasInput Measuring input 1, 2, diff. ditto
.Ipol Polarization current 0...1...±127 ditto
.Upol Polarization voltage 0...400...±1270 ditto
.PolElectrTest Test for polarized electrodes ON, OFF ditto
.Temp Titration temperature –170.0...25.0...500.0 ditto