Metrohm IC Net 2.0 Manuel d'utilisateur Page 53

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 196
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 52
5.4 761 Compact IC
IC Net 2.0
45
Second column Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the Remote configura-
tion tab.
Third column Parameter for program instruction (see
below).
ENABLED Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program subpage:
Instruction Parameter entry Meaning
Valve Inject, Fill Switch injection valve to "INJECT"
or "FILL" position.
FullScale 50, 250, 1000
µµS/cm
Set full scale range to the se-
lected value.
ICPump on, off Switch on or off the high-pressure
pump.
Flow 0.2 ... 2.5 mL/min Set flow rate of the high-pressure
pump to the desired value.
Pmax 0.0 ... 25.0 MPa Set maximum pressure limit for
the high-pressure pump to the de-
sired value.
Pmin 0.0 ... 25.0 MPa Set minimum pressure limit for
the high-pressure pump to the de-
sired value.
Remote 0, 1, *, p Set remote output lines 1...8 to
the desired values. For entry of the
first value, enter 1, 0, p or *. For en-
try of the other values, move the
cursor in front of the value to be
changed and enter 1, 0, p or *.
Vue de la page 52
1 2 ... 48 49 50 51 52 53 54 55 56 57 58 ... 195 196

Commentaires sur ces manuels

Pas de commentaire