Metrohm 797 VA Computrace Manuel d'utilisateur Page 127

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 295
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 126
5.3 Monitor
797 VA Computrace – Software
115
The window displays date and time of the deter-
mination start and the title of the method used.
Date and time in the format YYMMDDHHMM
(month-day-hour-minute) and the Sample ID are
used as default for the name of the determination
file to be saved automatically (e.g.
0706181712_sample.dth).
Sample ID [ 32 Zeichen ; "sample" ]
Identification for sample. The expression defined
in the Determination tab is displayed and can be
changed if desired.
Volume production bath (mL) [ > 0 mL ; 10 mL ]
The volume of sample bath solution added dis-
played and can be changed if desired.
Cell volume (mL) [ > 0 mL ; 10 mL ]
Total volume of solution in the measuring vessel at
the start of the determination. Is dependent on
whether you removed the electrolyte solution be-
fore you added the production bath or not.
Add the sample solution into the measuring vessel
and press <OK>.
PLACE INTERCEPT SOLUTION
This window appears in the Electroplating Bath VA
with the Calibration techniques “MLAT Standard
addition for brighteners” and "LAT Record calibra-
tion curve".
Vue de la page 126
1 2 ... 122 123 124 125 126 127 128 129 130 131 132 ... 294 295

Commentaires sur ces manuels

Pas de commentaire