Metrohm 797 VA Computrace Manuel d'utilisateur Page 239

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 295
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 238
8.6 Analyze Electroplating Bath Solutions
797 VA Computrace – Software
227
General Settings in the 797 Software
Settings on the Dosinos tab of the GENERAL SETTINGS window in
the 797 Software:
It is recommended to do for all used Dosinos one "Prep cycle".
On the Automation tab, select the 838 Advanced Sample Processor.
The default settings can be adopted (click button <Default>).
Method Parameters for the 797
Set CVS or CPVS for
Mode, and "MLAT" for Calibration.
Settings for the
DOSINOS window:
Checking Use for predose for Dosino 1, VMS and additional Sup-
pressor concentrate are added to the measuring vessel before the
sample transfer. Together, they form the "Intercept solution". To
keep the Suppressor concentration stable, additional Suppressor
concentrate is added after the sample transfer.
Vue de la page 238
1 2 ... 234 235 236 237 238 239 240 241 242 243 244 ... 294 295

Commentaires sur ces manuels

Pas de commentaire