Metrohm 797 VA Computrace Manuel d'utilisateur Page 253

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 295
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 252
8.6 Analyze Electroplating Bath Solutions
797 VA Computrace – Software
241
Example for a sample rack with 12 samples:
Rinsing solution
not used
Sample
1
2
3
28
27
29
30
31
56
55
57
58
59
84
83
85
86
87
112
111
Note: Although the first sample is placed on position 2, position
1 remains the starting position in the start sequence of the
method at the 838 (recording the Response curve).
Note: Before each start, define the position of the first sample
vessel for the 838 parameter "SAMPLE" with the 838 keypad.
General Settings in the 797 Software
Activate both options on the General tab of the GENERAL SET-
TINGS
window, so that a recorded calibration curve replaces the
old one automatically.
Settings on the Dosinos tab of the GENERAL SETTINGS window in
the 797 Software:
Vue de la page 252
1 2 ... 248 249 250 251 252 253 254 255 256 257 258 ... 294 295

Commentaires sur ces manuels

Pas de commentaire