Metrohm 797 VA Computrace Manuel d'utilisateur Page 58

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 295
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 57
3 General settings for exploratory and determination mode
797 VA Computrace – Software
46
NP – Normal Pulse (for "Exploratory" only)
General:
NP or Normal Pulse voltammetry is the classic pulse voltam-
metric measurement mode with direct recording of the current. It is
equally well suited for irreversible and reversible systems and offers
a higher sensitivity than the DC voltammetry. The NP measurement
mode can only be set for the exploratory mode by selecting NP -
Normal pulse
for the Mode parameter in the EXPLORATORY SPECI-
FICATION
window.
Description:
For NP voltammetry, square-wave pulses with an increasing ampli-
tude are superimposed on a constant base voltage. The current i is
measured as a function of the voltage U at the end of the pulse.
This normally provides wave-shaped curves which can be evaluated
using the tangent method.
U
Base potential
Voltage
step
Pulse
time
Voltage step time
t (i)t (i)
Start potential
Half-wave voltage
Wave current
i
U
Vue de la page 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 294 295

Commentaires sur ces manuels

Pas de commentaire