Metrohm IC Net 2.3 Manuel d'utilisateur Page 102

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 389
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 101
6 Devices
IC Net 2.3
92
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.
Hardware error
Hardware error detected at the 761 Com-
pact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
Alarm leak detector
Leak detector has detected solvent in the
instruments interior.
Remote line
Set remote output lines 18.
Selection: 0 (line off, inactive, open)
1 (line on, active, 0 V)
p (pulse output)
* (leave line in current status)
PowerOn values Display of the power on startup values for
remote output lines set on the Hardware tab.
Pulse length Length of a pulse in ms.
Alarm stops
The Alarm stops tab of the hardware settings window defines the
events for which the 761 Compact IC is stopped immediately. At
an alarm stop, high-pressure pump and peristaltic pump are
stopped immediately, the running determination and the active
sample queue are also stopped.
Vue de la page 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 388 389

Commentaires sur ces manuels

Pas de commentaire