7 Methods
IC Net 2.3
304
7.6.3 More items to report
This part of the Report options window contains a list of rarely
used report items that can be included into the report on the
user's choice:
Acquisition Parameters used for data acquisition from
the from the General, Measure and Filters
tabs of the Method setup window.
ACQUISITION PARAMETERS
Channels: 1
Method duration: 25.00min
Run duration: 25.01min
Measurements (method): 15006
Measurements (run): 15006
Freq.divisor: 1
Sampling: 10.00 pts/sec
Start delay: sec
Device:
Program before:
Program after:
Spikes filter: Yes
Median filter: No
slit: 0
Gauss filter: No
slit: 0
Integration Integration parameters and integration
events.
INTEGRATION DEFAULTS
Channel: ch1
Delay: 2.50 min
Width: 15.00 sec
Broadening: 2.00
Slope: 10.00
Asymmetry: 2.00
MinArea: 0.00
MinHeight: 1.00
Rider ratio: 0.00
No. min
1 5.00 Force horizontal base = 0.00
2 7.50 Cancel horizontal base = 0.00
Calibration defaults Calibration parameters defined in the com-
ponent window and on the Math tab of the
Method setup window.
CALIBRATION
Channel: ch1
Method: External standard
Response: Area
Standard: No
IDENTIFICATION
Reference peaks: Height
Other peaks: Time
Retention units: min