Metrohm IC Net 2.3 Manuel d'utilisateur Page 113

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 389
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 112
6.7 871 Advanced Bioscan
IC Net 2.3
103
opened by selecting this menu option with the right mouse button
or by double-clicking the 871 icon in the SYSTEM window. It con-
sists of the four tabs Manual, Method parameters, Program, and
Links.
Manual
This tab of the 871 Advanced Bioscan window allows manual con-
trol of the 871 Advanced Bioscan for a connected system. It dis-
plays the following current measurement values and parameter
settings of the 871 Advanced Bioscan:
Mode Selection of detector mode:
DC In DC mode, a constant potential is applied
to the working electrode, the analytes are
oxidized or reduced in accordance with
their electrochemical properties.
Pulse The Pulse mode works with three different
working potentials, which are applied cycli-
cally. This frees the electrode surface from
any adhering reaction products after each
sweep and an activated surface is pro-
duced for the next measurement.
Vue de la page 112
1 2 ... 108 109 110 111 112 113 114 115 116 117 118 ... 388 389

Commentaires sur ces manuels

Pas de commentaire