6 Devices
IC Net 2.3
110
E1 Start potential for sweep.
Range: -2 ... +2 V.
E2 End potential for sweep.
Range: -2 ... +2 V.
Rate Definition of scan rate in mV/s.
Selection: 1, 2, 5, 10, 20, 50 mV/s
Cycle mode Definition of scan cycle:
full double scan E1 ... E2 ... E1
half single scan E1 ... E2
Selection: full, half
Range
Selection of measuring range:
nA 10 ... 500 nA.
µA 1 ... 200 µA.
Offset Offset of the zero point for the analog out-
put signal.
Selection: -50 ... +50 % (in steps of 10 %)
Data rate Setting of the data acquisition rate.
Selection: -1, 2, 5, 10 Hz
Actual
E1, E2 Display of current start and end potentials
in V.
Rate Display of current scan rate in mV/s.
Cycle mode Display of current scan cycle mode.
Range Display of current measuring range.
Offset Display of current offset.
Data rate Display of current data rate.
<Set> Send current parameters immediately to the
871 Advanced Bioscan. Parameters are not
stored in the system file (*.smt) as long as
the file is not saved.
<Scan Window> Opens the 871 Scan shell window.