6 Devices
IC Net 2.3
76
Display
The Display tab of the 819 IC Detector window is only available for a
connected system. It displays the current measurement values
and parameters settings of the 819 IC Detector.
Conductivity Live display of measured conductivity
(auto-zero value) in µS/cm. This field is
available for the WATCH WINDOW. The
watch window color is changed to the Out
of range
color if the measured value is more
than 80% outside the
Full Scale range.
abs. Cond. Live display of absolute conductivity value
in µS/cm. This field is available for the
WATCH WINDOW. The watch window color is
changed to the
Out of range color if the
measured value is bigger or equal to 110%
of
Range.
Ion Display of ion settings:
Anion: polarity = +, temperature coeff. = 2.5
Cation: polarity = –, temperature coeff. = 1.5
Range Display of measuring range in µS/cm.
Full scale Display of full scale range in µS/cm.
Cell const. Display of cell constant of the conductivity
cell in /cm.
Thermostat Display of temperature within in the detec-
tor block.
Offset Display of offset of the zero point.