Metrohm IC Net 2.3 Manuel d'utilisateur Page 222

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 389
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 221
6 Devices
IC Net 2.3
212
tor P5 AUXILLIARIES. The time program
for each control line is arranged on a sepa-
rate sheet.
AUX-1,…AUX-4 Sheet for the time program of the corre-
sponding control line. For each control line
four events to set the contact closure to
normally open (Aux-X OFF) and four
events to set the contact closure to nor-
mally closed (Aux-X ON) can be pro-
grammed.
Current Displays the actual settings of the Timebase
program
.
Method values Check box to enable/disable the corre-
sponding event.
Disabled
Enabled A combo box appears to select
a time for the corresponding event.
Entry range: 0:00:00…9:59:59 h:mm:ss
<Set> Set parameter values selected under
Method values to the Timebase program.
Vue de la page 221
1 2 ... 217 218 219 220 221 222 223 224 225 226 227 ... 388 389

Commentaires sur ces manuels

Pas de commentaire