6 Devices
IC Net 2.3
104
Scan The Scan mode allows potential/current
curves to be recorded.
Cell Manual control of amperometric detector
cell:
<On> Switch on detector cell.
<Off> Switch off detector cell.
Oven Manual control of column oven (heater):
<On> Switch on oven.
<Off> Switch off oven.
Temperature Temperature of column oven.
Range: 15 ... 45 °C
Control Manual setting of 4 control lines.
<Set> Set control lines to selected states.
Aux1, Aux2, Cont1, Cont2
Setting of 4 control lines.
Selection: 0, 1, *
Valve Manual switching of the injection valve at
the 812 Valve Unit:
<Fill> Switch injection valve to "Fill" position (con-
trol line Cont1 is set to 0).
<Inject> Switch injection valve to "Inject" position
(control line Cont1 is set to 1).
Autozero
Manual setting of the analog output signal
to zero.
<Set> Turn on autozero function.
<Off> Turn off the autozero function.
Actual Display of actual values.
Mode Display of current detector mode.
Cell Display of current cell state.
Oven
Display of current ovent state.
Control
Display of current control line settings.
Valve
Display of current valve position.
Autozero Display of current autozero state.
Common
This tab of the 871 Advanced Bioscan window contains the method
startup values for the common parameters of the instrument.
These values are sent and applied to the 871 Advanced Bioscan
by Start determination and Startup hardware, or if the values are
sent manually with <Set>.