Metrohm IC Net 2.3 Manuel d'utilisateur Page 134

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 389
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 133
6 Devices
IC Net 2.3
124
Start with startup hardware
Start the Thermostat with the startup of the
system hardware.
Enabled The temperature of the column heater is
monitored by «IC Net», an injection is trig-
gered, if the set operating temperature of
the column heater is reached.
The operating temperature is reached,
when the actual temperature is within
±
1
°C of the temperature set above.
Disabled The temperature of the column heater is
not monitored by «IC Net». The proceed-
ing of a determination or a queue is not af-
fected if the set value is not reached.
Links
The Links tab of the 820 IC Separation Center is used for COM port
selection and settings (details see section 5.2.4 Links), resp. for
Event line selection.
Vue de la page 133
1 2 ... 129 130 131 132 133 134 135 136 137 138 139 ... 388 389

Commentaires sur ces manuels

Pas de commentaire